Abstract
Abstract
Purpose
To assess magnetic resonance imaging (MRI) artefacts caused by different computer-aided design/computer-aided manufacturing (CAD/CAM) retainers in comparison with conventional hand bent stainless steel twistflex retainers in vivo.
Materials and methods
MRI scans (3 Tesla) were performed on a male volunteer with different CAD/CAM retainers (cobalt–chromium, CoCr; nickel–titanium, NiTi; grade 5 titanium, Ti5) and twistflex retainers inserted. A total of 126 landmarks inside and outside the retainer area (RA; from canine to canine) were evaluated by two blinded radiologists using an established five-point visibility scoring (1: excellent, 2: good, 3: moderate, 4: poor, 5: not visible). Friedman and two-tailed Wilcoxon tests were used for statistical analysis (significance level: p < 0.05).
Results
Twistflex retainers had the strongest impact on the visibility of all landmarks inside (4.0 ± 1.5) and outside the RA (1.7 ± 1.2). In contrast, artefacts caused by CAD/CAM retainers were limited to the dental area inside the RA (CoCr: 2.2 ± 1.2) or did not impair MRI-based diagnostics in a clinically relevant way (NiTi: 1.0 ± 0.1; Ti5: 1.4 ± 0.6).
Conclusion
The present study on a single test person demonstrates that conventional stainless steel twistflex retainers can severely impair the diagnostic value in head/neck and dental MRI. By contrast, CoCr CAD/CAM retainers can cause artefacts which only slightly impair dental MRI but not head/neck MRI, whereas NiTi and Ti5 CAD/CAM might be fully compatible with both head/neck and dental MRI.
Funder
Medizinische Fakultät Heidelberg der Universität Heidelberg
Publisher
Springer Science and Business Media LLC
Subject
Oral Surgery,Orthodontics
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献