Multibraided Fixed Retainers with Different Diameters after Magnetic Resonance Imaging (MRI): In Vitro Study Investigating Temperature Changes and Bonding Efficacy

Author:

Sfondrini Maria Francesca1ORCID,Pascadopoli Maurizio1ORCID,Gandini Paola1,Preda Lorenzo234ORCID,Sfondrini Domenico5,Bertino Karin1,Rizzi Cinzia1,Scribante Andrea16ORCID

Affiliation:

1. Unit of Orthodontics and Pediatric Dentistry, Section of Dentistry, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy

2. Diagnostic Imaging and Radiotherapy Unit, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy

3. Department of Radiology, Fondazione IRCCS Policlinico San Matteo, University of Pavia, 27100 Pavia, Italy

4. National Center for Oncological Hadrontherapy (CNAO), 27100 Pavia, Italy

5. Maxillo-Facial Surgery Unit, Fondazione IRCCS Policlinico San Matteo, 27100 Pavia, Italy

6. Unit of Dental Hygiene, Section of Dentistry, Department of Clinical, Surgical, Diagnostic and Pediatric Sciences, University of Pavia, 27100 Pavia, Italy

Abstract

Objectives: Orthodontists are often asked to remove fixed retainers before patients undergo magnetic resonance imaging (MRI). The present in vitro study was designed to analyze the heating and bonding efficacy of stainless steel multibraided fixed retainers after 1.5- and 3-tesla (T) MRI. Materials and methods: A total of 180 human mandibular incisors were used to create 45 specimens of four teeth each, divided into nine groups. Handmade multibraided fixed retainers of three different sizes, defined by the diameter of the initial wire used (0.008″, 0.010″ and 0.012″), were tested. Three groups underwent MRI at 1.5 T, another three groups underwent MRI at 3 T and the last three groups did not undergo MRI. Temperature was assessed before and after MRI. Shear bond strength (SBS) and adhesive remnant index (ARI) were assessed after MRI for all groups. Data were statistically analyzed (p < 0.05). Results: After 1.5 T exposure, no significant temperature increase from T0 to T1 was observed in any of the groups (p > 0.05). Regarding the 3 T groups, a significant difference from T0 to T1 was found for all the groups (p < 0.05). Temperature changes were not clinically relevant, as they were less than 1 °C for all groups except for group 3 (ΔT0–T1: 1.18 ± 0.3 °C) and group 6 (ΔT0–T1: 1.12 ± 0.37 °C). Furthermore, there were no significant differences between the temperature variations associated with different wire diameters (p > 0.05). Conclusions: No significant changes in SBS or ARI were found (p > 0.05). Clinical significance: Since overheating was irrelevant and adhesion values did not change, the tested devices were concluded to be safe for MRI examinations at 1.5 T and 3 T.

Publisher

MDPI AG

Reference40 articles.

1. Influence of common orthodontic appliances on the diagnostic quality of cranial magnetic resonance images;Elison;Am. J. Orthod. Dentofac. Orthop.,2008

2. Effects of orthodontic appliances on the diagnostic capability of magnetic resonance imaging in the head and neck region: A systematic review;Hasanin;Int. Orthod.,2019

3. Unexpected complications associated with mandibular fixed retainers: A retrospective study;Kucera;Am. J. Orthod. Dentofac. Orthop.,2016

4. Orthodontic retreatment need and related self-perceived factors among contemporary college freshmen in China;Wang;J. Clin. Pediatr. Dent.,2023

5. Magnetic resonance imaging diagnosis of the temporomandibular joint in patients with orthodontic appliances;Okano;Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endod.,2003

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3