Design of an Experimental Test System to Investigate Parameters Affecting Distal Tip Loads of Pacemaker and Defibrillator Leads

Author:

Stephen Elizabeth A.1,Walsh Donna L.2,Duraiswamy Nandini1,Vesnovsky Oleg1,Timmie Topoleski L. D.13

Affiliation:

1. U.S. Food and Drug Administration, Center for Devices and Radiological Health, 10903 New Hampshire Avenue, Silver Spring, MD 20993

2. U.S. Food and Drug Administration, Center for Devices and Radiological Health, 10903 New Hampshire Avenue, Silver Spring, MD 20993 e-mail:

3. Department of Mechanical Engineering, University of Maryland–Baltimore County, 1000 Hilltop Circle, Baltimore, MD 21250

Abstract

The purpose of this study was to design and evaluate a system to test the mechanical behavior of pacemaker and defibrillator leads. Over 300,000 pacemaker and implantable cardioverter defibrillator (ICD) procedures are performed every year in the U.S. for the treatment of cardiac arrhythmias, ventricular dysrhythmias, and congestive heart failure. These procedures require implanting transvenous leads into the interior wall of the heart. A serious and sometimes fatal complication that may occur during or after lead implantation is perforation of the lead tip through the heart wall. The factors that lead to perforation are not fully understood. This illustrates that the mechanical interactions between the lead tip and the cardiac tissue need to be further investigated to improve the outcome for pacemaker and ICD patients. To improve the performance of lead tips, the testing protocols must reproduce physiological and clinically relevant tip-tissue interactions. As a first step toward this goal, testing parameters that influence those interactions must be identified. We investigated the effect of test system parameters, which reproduce potentially critical physiological constraints, on the load experienced at the distal tip of thirteen pacemaker and defibrillator active-fixation leads. We evaluated the use of a constraint to simulate the effect of the right ventricle (RV constraint) in vivo, how and where the lead was fixed in the test configuration, location of the load cell in the test system, rotation and frequency of the test protocol, and the effect of stylets. Results showed the RV constraint and load cell placement had the largest impact on lead tip load, while rotation of the test setup and test frequency had a minimal impact. Recommendations are made for a test system and protocol for in vitro testing of leads that take into consideration in vivo conditions. Better approximations of the in vivo environment may lead to improved product development. The potential of this system to more effectively evaluate new pacemaker and defibrillator lead designs will require further study.

Publisher

ASME International

Subject

Biomedical Engineering,Medicine (miscellaneous)

Reference19 articles.

1. Prevalence and Characterization of Asymptomatic Pacemaker and ICD Lead Perforation on CT;PACE,2007

2. Heart Disease and Stroke Statistics—2013 Update: A Report From the American Heart Association;American Heart Association Statistics Committee and Stroke Statistics Committee;Circulation,2013

3. The 11th World Survey of Cardiac Pacing and Implantable Cardioverter-Defibrillators: Calendar Year 2009—A World Society of Arrhythmia's Project;PACE,2011

4. Permanent Pacemaker Implantation Technique: Part II;Heart,2009

5. Incidence and Predictors of Cardiac Perforation After Permanent Pacemaker Placement;Heart Rhythm,2005

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3