Optimization of Electrode Configuration and Pulse Strength in Irreversible Electroporation for Large Ablation Volumes Without Thermal Damage

Author:

Yang Yongji1,Moser Michael2,Zhang Edwin3,Zhang Wenjun4,Zhang Bing5

Affiliation:

1. Tumor Ablation Group, CISR Center, East China University of Science and Technology, Shanghai 200237, China e-mail:

2. Department of Surgery, University of Saskatchewan, Saskatoon, SK S7N 0W8, Canada e-mail:

3. Division of Vascular and Interventional Radiology, Department of Medical Imaging, University of Toronto, Toronto, ON M5T 1W7, Canada e-mail:

4. Fellow ASME Tumor Ablation Group, CISR Center, East China University of Science and Technology, Shanghai 200237, China e-mail:

5. Mem. ASME School of Mechatronic Engineering and Automation, Shanghai University, Shanghai 200072, China e-mail:

Abstract

The aim of this study was to analyze five factors that are responsible for the ablation volume and maximum temperature during the procedure of irreversible electroporation (IRE). The five factors used in this study were the pulse strength (U), the electrode diameter (B), the distance between the electrode and the center (D), the electrode length (L), and the number of electrodes (N). A validated finite element model (FEM) of IRE was built to collect the data of the ablation volume and maximum temperature generated in a liver tissue. Twenty-five experiments were performed, in which the ablation volume and maximum temperature were taken as response variables. The five factors with ranges were analyzed to investigate their impacts on the ablation volume and maximum temperature, respectively, using analysis of variance. Response surface method (RSM) was used to optimize the five factors for the maximum ablation volume without thermal damage (the maximum temperature ≤ 50 °C for 90 s). U and L were found with significant impacts on the ablation volume (P < 0.001, and P = 0.009, respectively) while the same conclusion was not found for B, D and N (P = 0.886, P = 0.075 and P = 0.279, respectively). Furthermore, U, D, and N had the significant impacts on the maximum temperature with P < 0.001, P < 0.001, and P = 0.003, respectively, while same conclusion was not found for B and L (P = 0.720 and P = 0.051, respectively). The maximum ablation volume of 2952.9960 mm3 without thermal damage can be obtained by using the following set of factors: U = 2362.2384 V, B = 1.4889 mm, D = 7 mm, L = 4.5659 mm, and N = 3. The study concludes that both B and N have insignificant impacts (P = 0.886, and P = 0.279, respectively) on the ablation volume; U has the most significant impact (P < 0.001) on the ablation volume; electrode configuration and pulse strength in IRE can be optimized for the maximum ablation volume without thermal damage using RSM.

Publisher

ASME International

Subject

General Earth and Planetary Sciences,General Environmental Science

Reference33 articles.

1. Membrane Electroporation Theories: A Review;Med. Biol. Eng. Comput.,2006

2. Electroporation

3. Electroporation of Cells and Tissues;IEEE Trans. Plasma Sci.,2000

4. Irreversible Electroporation in Medicine;Technol. Cancer Res. Treat.,2007

5. Intracranial Nonthermal Irreversible Electroporation: In Vivo Analysis;J. Membr. Biol.,2010

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3