Computational Fluid Dynamics-Based Design Optimization for an Implantable Miniature Maglev Pediatric Ventricular Assist Device

Author:

Wu Jingchun1,Antaki James F.2,Verkaik Josiah3,Snyder Shaun3,Ricci Michael3

Affiliation:

1. LaunchPoint Technologies, Inc., 5735 Hollister Ave., Suite B, Goleta, CA 93117

2. Department of Biomedical Engineering and Computer Science, Carnegie Mellon University, 700 Technology Drive, Suite 4321, Pittsburgh, PA 15213

3. Launch Point Technologies, Inc., 5735 Hollister Avenue, Suite B, Goleta, CA 93117

Abstract

Computational fluid dynamics (CFD)-based design optimization was applied to achieve the finalized design of the PediaFlow® PF4, a magnetically levitated rotodynamic pediatric ventricular assist device. It features a streamlined blood-flow path with a single annular fluid passage between the rotor and the stationary housing. The resulting impeller is composed of a first-stage mixed-flow section having four blades at the conical nose region followed by a second-stage fully axial-flow section with three blades within the annular gap region. A stator with three inwardly-directed vanes is provided at the conical tail region to recover pressure and straighten the flow. CFD predictions of head and efficiency characteristics agreed remarkably well with the validation experimental data: with overprediction of head by <7 mmHg over the entire operational range and a slight overprediction in best efficiency by ∼1%. The new optimized PF4 extended the maximum flow range of the previous PF3 device by more than 100% to over 2.3 liter per minute (LPM) for the same range of operating speeds, and doubled the maximum hydraulic efficiency to ∼27%. Evaluation of hemolysis was performed by a Lagrangian particle-tracking technique with analysis of regional contributions to the overall blood damage. The simulation revealed that hemolysis increases with an increase in both the flow rate and rotor speed but not necessarily with just an increase in flow rate at a constant rotor speed. At the flow rate of 1.0 LPM and a head of 138 mmHg, PF4 has a hemolysis index of 0.0032 compared to 0.0058 produced by PF3 at the same flow rate with a head of 48 mmHg. Numerical simulation of radial fluid forces performed by the CFD model with an eccentric rotor revealed the presence of negative fluid stiffness that was monotonically related to both flow and speed. Finally, conjugate heat transfer analysis predicted temperature rise adjacent to the motor to be inversely proportional to the length, but not exceeding ∼2 °C over the intended range of operation. In conclusion, CFD-based design optimization greatly expedited and facilitated the completion of the PediaFlow® flow path and contributed to the system-wide optimization to produce a miniature maglev pump with exceptional hemocompatibility.

Publisher

ASME International

Subject

Mechanical Engineering

Cited by 21 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3