Analysis of Factors Influencing Accuracy of Volume Flow Measurement in Dialysis Access Fistulas Based on Duplex Ultrasound Simulation

Author:

Zierler R. Eugene1ORCID,Leotta Daniel F.2,Sansom Kurt3,Aliseda Alberto3,Anderson Mark D.4,Sheehan Florence H.4

Affiliation:

1. Division of Vascular Surgery, Department of Surgery, University of Washington, Seattle, WA, USA

2. Applied Physics Laboratory, University of Washington, Seattle, WA, USA

3. Department of Mechanical Engineering, University of Washington, Seattle, WA, USA

4. Division of Cardiology, Department of Medicine, University of Washington, Seattle, WA, USA

Abstract

Objective:We developed a duplex ultrasound simulator and used it to assess accuracy of volume flow measurements in dialysis access fistula (DAF) models.Methods:The simulator consists of a mannequin, computer, and mock transducer. Each case is built from a patient’s B-mode images that are used to create a 3-dimensional surface model of the DAF. Computational fluid dynamics is used to determine blood flow velocities based on model vessel geometry. The simulator displays real-time B-mode and color-flow images, and Doppler spectral waveforms are generated according to user-defined settings. Accuracy was assessed by scanning each case and measuring volume flow in the inflow artery and outflow vein for comparison with true volume flow values.Results:Four examiners made 96 volume flow measurements on four DAF models. Measured volume flow deviated from the true value by 35 ± 36%. Mean absolute deviation from true volume flow was lower for arteries than veins (22 ± 19%, N = 48 vs. 58 ± 33%, N = 48, p < 0.0001). This finding is attributed to eccentricity of outflow veins which resulted in underestimating true cross-sectional area. Regression analysis indicated that error in measuring cross-sectional area was a predictor of error in volume flow measurement (β = 0.948, p < 0.001). Volume flow error was reduced from 35 ± 36% to 9 ± 8% ( p < 0.000001) by calculating vessel area as an ellipse.Conclusions:Duplex volume flow measurements are based on a circular vessel shape. DAF inflow arteries are circular, but outflow veins can be elliptical. Simulation-based analysis showed that error in measuring volume flow is mainly due to assumption of a circular vessel.

Funder

National Institute of Biomedical Imaging and Bioengineering

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,General Medicine,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3