Abstract

A survey of [H+], Pco2, and Po2 analyses in 360 laboratories was conducted using three commercial control materials: two aqueous gas-equilibrated buffer solutions (General Diagnostics and IL) and one whole blood material (DADE). There was little difference in precision or accuracy between instruments, classified according to manufacturer into four groups, and materials for [H+] and Pco2, or in precision for Po2. There were, however, differences in accuracy between instrument groups for Po2 analysis on the aqueous materials, and in some cases the mean values lay outside the range assigned by the material manufacturers. The 35 instruments outside clinical chemistry departments yielded results similar to those of all participants. The relation between inter- and intra-laboratory precision was similar to that for many analytes, suggesting that neither these materials nor the state of these analyses is unsatisfactory. Aqueous materials should, however, be used with caution as accuracy controls for Po2.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Regional Quality Assessment of pH and Blood Gas Analysers;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;1986-01

2. Monthly Inter laboratory pH and Blood-gas Survey: Establishing Accuracy Based on Inter laboratory Performance;American Journal of Clinical Pathology;1984-02-01

3. A Review of Blood pH and Blood-Gas Analysis;Annals of Clinical Biochemistry: International Journal of Laboratory Medicine;1982-07

4. Inter-laboratory comparison of acid-base variables in human blood and in quality control materials;Clinica Chimica Acta;1981-11

5. Oxygen Tension Measurement Using an Automatic Blood Gas Analyser;Journal of the Royal Society of Medicine;1981-08

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3