Affiliation:
1. Department of Biochemical Medicine, Stracathro Hospital, Brechin
Abstract
Summary pH, Pco2, and Po2 measurements have been made over a three-week period on General Diagnostics Blood GAS Control Ampoules and on a tonometered buffer system. The latter seems adequate as a precision and accuracy control of pH, Pco2, and Po2 measurements in a single solution. The former seems adequate as a precision control of these measurements, but, on the particular batch provided, mean values need to be determined within the customer's laboratory.
Subject
Clinical Biochemistry,General Medicine
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献