Affiliation:
1. Department of Chemical Pathology, St George's Hospital Medical School, London SW170RE, UK
Abstract
Data obtained from routine analytical radioimmunoassays were processed using five curve-fitting procedures, viz. ‘Amersham’, single binding site, four parameter logistic, a linear logit-log and a polynomial logit-log. The polynomial logit-log procedure gave the best fit, but this was probably due to the inherent flexibility of this curve-fitting process since the analytical precision achieved with it was no better than what was obtained with most of the other procedures. A limited study failed to show that statistical weighting of data before curve fitting had any practical advantage.
Subject
Clinical Biochemistry,General Medicine
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献
1. The effect of the immunoassay curve fitting routine on bias in troponin;Clinical Chemistry and Laboratory Medicine (CCLM);2022-10-26
2. Calibration Curve Fitting;The Immunoassay Handbook;2013
3. 7. Bibliography;Scandinavian Journal of Clinical and Laboratory Investigation;1987-01