Background Biotin interference in streptavidin-based immunoassays causes widespread analytical distortions that may lead to clinical confusion, inappropriate patient management and, ultimately, adverse events. Its prevalence has increased recently due to the increased use of high-dose biotin therapy in specific patient groups (notably multiple sclerosis) and possibly the general community. Methods We have developed a method to deplete biotin from samples using the streptavidin-coated magnetic microparticles that are a component of most susceptible assays. Results We show that high concentrations of spiked biotin can be adequately depleted from serum using this approach, and that gross biochemical derangements can be restored to normality. We also show that biotin in samples derived from multiple sclerosis patients receiving 300 mg biotin daily can be adequately depleted to remove associated analytical interference and restore normal results. The method is applicable to competitive and sandwich immunoassays and importantly, because it does not change the volume of the sample, suitable for the measurement of free thyroid hormone assays. Application of the method does not significantly change the precision of measurement, and for the majority of analytes, the accuracy is not substantially altered. Conclusions Adopting this method enables laboratories to confirm biotin interference in the appropriate clinical setting. Moreover, it enables laboratories to remove the interference and report accurate and reliable results, without the need for patients to withhold beneficial therapy prior to blood tests. Until the biotin tolerance of susceptible assays is improved, our method gives laboratories a safe alternative for reporting results using streptavidin-based methods.

Publisher

SAGE Publications

Subject

Clinical Biochemistry,General Medicine

Cited by 62 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3