The automated laser-based hematology analyzer Sysmex XT-2000 iV™ provides a 5-part differential count and specific cytograms that are of great interest for large veterinary laboratories. The aim of the study was to validate the Sysmex XT-2000 iV compared to the laser-based hematology analyzer ADVIA® 2120 and manual differential in dogs, cats, and horses as well as the impact of anticoagulant (heparin, ethylenediamine tetra-acetic acid [EDTA], and citrate) and storage at 22°C and 4°C. Consecutive fresh K3–EDTA blood samples from 216 cats, 314 dogs, and 174 horses were included. The impact of anticoagulant and sample storage was assessed in specimens obtained from an additional 9 cats, 10 dogs, and 10 horses. Agreement between both analyzers was excellent to good except for monocytes and canine reticulocytes. Spearman rank correlation coefficients ( rs) between Sysmex XT-2000 iV and manual differential were good to fair and ranged from 0.91 (cat lymphocytes) to 0.44 (cat monocytes). Hematocrit value (Hct), mean corpuscular hemoglobin (MCH), MCH concentration (MCHC; all: P < 0.001), and mean corpuscular volume (MCV; P < 0.01) were higher in canine citrated blood compared to heparin and EDTA. In cats, lymphocytes and monocytes were lower in heparinized blood compared to EDTA ( P < 0.05), whereas in horses no significant effect was seen. Regarding storage time and temperature, white and red blood cell counts, hemoglobin, and MCH were stable. Hct, MCV, and MCHC were influenced by erythrocyte swelling. Differential count remained stable for 24 hr (22°C) and nearly 72 hr (4°C) except for monocytes. The overall performance of the Sysmex XT-2000 iV was excellent and compared favorably with that of the ADVIA 2120. A special strength was the excellent detection of feline eosinophils.

Publisher

SAGE Publications

Subject

General Veterinary

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3