Abstract
Background Femoral head collapse and coxa vara lead to internal fixator failure in elderly patients with hip fracture. External fixator application is an optimal choice; however, the existing methods have many disadvantages. Methods Type 31-A1.3 hip fracture models were developed in nine pairs of 1-year-old fresh bovine corpse femur specimens. Each left femur specimen was fixed by a dynamic hip screw (control group), and each right femur specimen was fixed by the slide-poking external fixator (experimental group). Vertical loading and torsion tests were then performed in both groups. Results In the vertical loading experiment, a 1000-N load was implemented. The mean vertical downward displacement of the femoral head in the experimental and control groups was 1.49322 ± 0.116280 and 2.13656 ± 0.166374 mm, respectively. In the torsion experiment, when the torsion was increased to 10.0 Nm, the mean torsion angle in the experimental and control groups was 7.9733° ± 1.65704° and 15.4889° ± 0.73228°, respectively. The slide-poking external fixator was significantly more resistant to compression and rotation than the dynamic hip screw. Conclusion The slide-poking external fixator for hip fractures that was designed and developed in this study can provide sufficient stability to resist compression and rotation in hip fractures.
Subject
Biochemistry, medical,Cell Biology,Biochemistry,General Medicine