Trifecta versus Perimount Magna Ease aortic valves: Failure mechanisms

Author:

Suzuki Ryo1ORCID,Ito Toshiro1,Suzuki Masato1,Ohori Shunsuke1,Takayanagi Ryo1,Miura Shiro1

Affiliation:

1. Department of Cardiovascular Surgery, Hokkaido Ohno Memorial Hospital, Hokkaido, Japan

Abstract

Background There are increasing reports of early externally mounted pericardial Trifecta bioprosthesis failure. We compared the hemodynamic performance of Trifecta and Carpentier–Edwards Perimount Magna Ease valves to determine the failure mechanism. Methods We retrospectively included 270 consecutive patients (age: 73.4 ± 8.2 years; 57.5% male; mean follow-up: 48.0 ± 20.3 months) who underwent aortic valve replacement from 2014 to 2021 at a single center and compared the Trifecta ( N = 137) and Carpentier–Edwards Perimount Magna Ease valve ( N = 133) patients. Results The prosthetic valve major aortic regurgitation incidence was higher for the Trifecta than that for the Carpentier–Edwards Perimount Magna Ease valve (6.3% vs. 0%, P < 0.009). Among the Trifecta failures, 33% developed structural valve deterioration, but all requiring redo aortic valve replacement developed major prosthetic valve aortic regurgitation. Freedom at 5 years from redo aortic valve replacement due to structural valve deterioration was significantly lower for Trifecta (89.4% vs. 100%, P = 0.003). The reoperation hazards were determined for Trifecta (vs. Carpentier–Edwards Perimount Magna Ease): 11.6 (1.47–90.9; P = 0.02), prosthetic valve aortic regurgitation: 2.38 (1.70–3.32; P < 0.01), structural valve deterioration: 20.82 (4.08–106.2; P < 0.01), 5-year mean transprosthetic pressure gradient: 1.14 per 1-point increase (1.03–1.24; P = 0.007), and urgent surgery: 10.1 (2.59–39.0; P = 0.001). The Cox regression analysis identified that prosthetic valve aortic regurgitation solely contributed to redo aortic valve replacement (hazard ratio: 2.38; confidence intervals: 1.70–3.32). Conclusions Significantly, more early failures occurred with the Trifecta valve than the Carpentier–Edwards Perimount Magna Ease valve but the Trifecta showed reasonable mean transprosthetic pressure gradient over time. Prosthetic valve aortic regurgitation and calcific structural valve deterioration synergistically contributed to Trifecta valve failure alternatively.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Pulmonary and Respiratory Medicine,General Medicine,Surgery

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3