Clinical evaluation of nine hollow-fibre membrane oxygenators

Author:

Segers P AM1,Heida J F2,de Vries I2,Maas C2,Boogaart A J2,Eilander S2

Affiliation:

1. Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle,

2. Department of Extracorporeal Circulation, Isala Clinics, Weezenlande Hospital, Zwolle

Abstract

In a comparative study we investigated the performance characteristics of nine hollow-fibre oxygenators. In a clinical setting, 10 units of each type of oxygenator were tested for oxygen exchange, transoxygenator pressure drop, heat exchanger performance and blood trauma. The oxygenators included are Maxima PRF Plus, Affinity, Forté, Affinity NT, Quantum, Optima, Capiox 1.8, Hilite and Quadrox. Ninety patients scheduled for elective coronary bypass surgery were enrolled in the study. Cardiotomy suction blood was not allowed to mix with the circulating blood. Samples were taken preoperatively (t1), immediately after cooling and crossclamping of the aorta (t2), during rewarming (t3), and postoperatively (t4). The results showed that under clinical conditions all of the tested oxygenators are well capable of meeting the normal and above-normal oxygen needs. The Quantum and Affinity NT are the best performing oxygenators with respect to oxygen transfer. The Quantum oxygenator with a small membrane surface area of 1.9 m2 has also the highest oxygen transfer per square metre. A direct relationship between shunt fraction and blood flow could not be established in all oxygenators. The lowest shunt fraction at 4 litres blood flow was seen in the Maxima, Quantum and Affinity NT groups (15%). At near physiological PaO2-values, we noticed a negative influence on the shunt fraction. Large differences exist in pressure drop between the oxygenators under standardized conditions with circulating priming, as well as during bypass. The lowest pressure drop was noticed in the Quadrox and Affinity NT, the highest in the Quantum, Capiox and Optima. However, we did not measure any differences in pressure drop after cooling and during re-warming. Differences exist between the heat exchanger performances. The Quadrox oxygenator is the most efficient, while the Maxima has the smallest performance factor. It was not possible to calculate heat exchanger performance for the Hilite due to incorrect temperature measurement. The high pressure gradients in some of the devices did not result in higher haemolysis. Plasma free haemoglobin values were low in each oxygenator. There are no differences in platelet drop postoperatively. The influence on blood trauma of the higher pressure drop in some of the tested devices, in combination with the higher centrifugal pump revolutions needed to overcome this gradient, has to be studied with longer perfusion times.

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 43 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3