Reduced embolic load during clinical cardiopulmonary bypass using a 20 micron arterial filter

Author:

Jabur GNS1,Willcox TW12,Zahidani SH1,Sidhu K1,Mitchell SJ12

Affiliation:

1. Green Lane Clinical Perfusion, Auckland City Hospital, Auckland, New Zealand

2. Department of Anesthesiology, Faculty of Medical and Health Sciences, University of Auckland, Auckland, New Zealand

Abstract

Objective: To compare the efficiency of 20 and 40 µm arterial line filters during cardiopulmonary bypass for the removal of emboli from the extracorporeal circuit. Methods: Twenty-four adult patients undergoing surgery were perfused using a cardiopulmonary bypass circuit containing either a 20 µm or 40 µm arterial filter (n = 12 in both groups). The Emboli Detection and Classification system was used to count emboli upstream and downstream of the filter throughout cardiopulmonary bypass. The mean proportion of emboli removed by the filter was compared between the groups. Results: The 20 µm filter removed a significantly greater proportion of incoming emboli (0.621) than the 40 µm filter (0.334) (p=0.029). The superiority of the 20 µm filter persisted across all size groups of emboli larger than the pore size of the 40 µm filter. Conclusion: The 20 µm filter removed substantially more emboli than the 40 µm filter during cardiopulmonary bypass in this comparison.

Publisher

SAGE Publications

Subject

Advanced and Specialized Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology, Nuclear Medicine and imaging,General Medicine

Cited by 20 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3