Clinical evaluation of the oxygenation capacity and controllability of 15 commercially available membrane oxygenators during alpha-stat regulated hypothermic cardiopulmonary bypass

Author:

Stinkens D.1,Himpe D.2,Thyssen P.1,De Bakker A.1,Smets W.1,Borms S.2,Muylaert Ph3,Van Hove M.3,Suy M.2,Van Cauwelaert Ph4

Affiliation:

1. Extracorporeal Circulation Technology Department, AZ Middelheim

2. Department of Anaesthesiology, AZ Middelheim

3. Department of Cardiovascular Surgery, W Theunissen Department of Anaesthesiology, AZ Middelheim

4. Department of Cardiovascular Surgery, AZ Middelheim, Lindendreef 1, 2020 Antwerp

Abstract

Oxygenation performance was tested in 15 membrane oxygenators by calculating the resistance for oxygenation (R) and the calculation of oxygen transferred versus FiO2. The clinical data reveal that the SARNS Turbo 440 (36.7-51.9%), the AFFINITY (37.2-50.1 %) and the HF 5400 (37.5-52.3%) are the oxygenators with the lowest FiO2 settings for comparable amounts of oxygen transferred during hypothermia and during normothermia followed by MAXIMA Plus (39.1-55.8%), MAXIMA Plus PRF (39.1-56.2%), CAPIOX SX 18 (39.7-61.2%), MONOLYTH (43.0-61.3%), OXIM 11-34 (44.1-63.9%), COBE Duo (44.7-64.9%), COBE Optima (47.4-66.4%), COMPACTFLO (48.3-65.3%), SAFE II (49.0-67.6%), UNIVOX (49.8-71.3%), MAXIMA (50.2-70.1 %) and the CM 50 (58.6-77.0%). Similar results were found by calculation of R. HF 5400 (2.41-1.87 mmHg/min/ml O2), AFFINITY (2.63-1.87 mmHg/min/ml O2), OXIM II-34 (2.72-2.45 mmHg/min/ml O2), MAXIMA Plus PRF (2.75-2.07 mmHg/min/ ml O2), COBE Optima (2.83-2.13 mmHg/min/ml O2), UNIVOX (2.84-2.17 mmHg/min/ml O2), MONOLYTH (2.89-2.24 mmHg/min/ml O2), SARNS Turbo 440 (3.03-2.12 mmHg/min/ml O2), MAXIMA Plus (3.09-2.28 mmHg/min/ml O2), SAFE 11 (3.19-2.50 mmHg/min/ml O2), CAPIOX SX 18 (3.27-2.44 mmHg/ min/ml O2), COMPACTFLO (3.41-2.50 mmHg/min/ml O2), MAXIMA (3.53-2.72 mmHg/min/ml O2), COBE Duo (3.57-2.71 mmHg/min/ml O2) and the CM 50 (3.53-2.72 mmHg/min/ml O2). As a measure of controllability of oxygenation, the coefficient of variation on the FiO2 ordered on the normothermic blood samples was used, giving as a result CAPIOX SX 18 (2.9-2.5%), AFFINITY (3.6-3.5%), COBE Duo (4.3-2.9%), HF 5400 (5.7-4.7%), MAXIMA Plus (8.1- 5.4%), COMPACTFLO (8.2-5.0%), MONOLYTH (8.3-4.0%), MAXIMA (8.7- 3.4%), COBE Optima (9.6-6.8%), SARNS Turbo 440 (10.1-7.3%), MAXIMA Plus PRF (10.9-8.7%), CM 50 (11.9-2.4%), UNIVOX (13.3-8.9%), OXIM 11-34 (15.5-17.3%) and the SAFE II (16.1-9.8%). The low FiO2 settings and the lower resistance for oxygenation are an indication of the reserve capacity of the oxygenators whose importance is proven by clinical data of emergency perfusions on patients under full resuscitation.

Publisher

SAGE Publications

Subject

Advanced and Specialised Nursing,Cardiology and Cardiovascular Medicine,Safety Research,Radiology Nuclear Medicine and imaging,General Medicine

Cited by 6 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3