Iterative reconstruction with multifrequency signal recognition technology to improve low-contrast detectability: A phantom study

Author:

Funama Yoshinori1ORCID,Shirasaka Takashi23,Goto Taiga4,Aoki Yuko4,Tanaka Kana4,Yoshida Ryo4

Affiliation:

1. Department of Medical Radiation Sciences, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan

2. Graduate School of Health Sciences, Kumamoto University, Kumamoto, Japan

3. Division of Radiology, Department of Medical Technology, Kyushu University, Fukuoka, Japan

4. Rad Diagnostic R&D Division, Radiation Diagnostic Systems Division, FUJIFILM Healthcare Corporation, Tokyo, Japan

Abstract

Background Brain CT needs more attention to improve the extremely low image contrast and image texture. Purpose To evaluate the performance of iterative progressive reconstruction with visual modeling (IPV) for the improvement of low-contrast detectability (IPV-LCD) compared with filtered backprojection (FBP) and conventional IPV. Materials and methods Low-contrast and water phantoms were used. Helical scans were conducted with the use of a CT scanner with 64 detectors. The tube voltage was set at 120 kVp; the tube current was adjusted from 60 to 300 mA with a slice thickness of 0.625 mm and from 20 to 150 mA with a slice thickness of 5.0 mm. Images were reconstructed with the FBP, conventional IPV, and IPV-LCD algorithms. The channelized Hotelling observer (CHO) model was applied in conjunction with the use of low-contrast modules in the low-contrast phantom. The noise power spectrum (NPS) and normalized NPS were calculated. Results At the same standard and strong levels, the IPV-LCD method improved low-contrast detectability compared with the conventional IPV, regardless of contrast-rod diameters. The mean CHO values at a slice thickness of 0.625 mm were 1.83, 3.28, 4.40, 4.53, and 5.27 for FBP, IPV STD, IPV-LCD STD, IPV STR, and IPV-LCD STR, respectively. The normalized NPS for the IPV-LCD STD and STR images were slightly shifted to the higher frequency compared with that for the FBP image. Conclusion IPV-LCD images further improve the low-contrast detectability compared with FBP and conventional IPV images while maintaining similar FBP image appearances.

Funder

FUJIFILM Healthcare Corporation

Publisher

SAGE Publications

Subject

General Earth and Planetary Sciences,General Environmental Science

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3