In vivo comparison of implantation behavior and laboratory analysis of two preloaded intraocular lens injectors

Author:

Baur Isabella D1ORCID,Łabuz Grzegorz1,Chychko Lizaveta1ORCID,Yildirim Timur M1,Naujokaitis Tadas1,Auffarth Gerd U1ORCID,Khoramnia Ramin

Affiliation:

1. David J. Apple Center for Vision Research, Department of Ophthalmology, University of Heidelberg, Heidelberg, Germany

Abstract

Purpose We evaluated implantation behavior and injector damage of two different IOL injector systems, the Multisert and the AutonoMe. Design Prospective, randomized, comparative study with laboratory investigation. Methods We examined used injectors from 48 bilateral cataract cases and assessed video recordings of each Implantation. All eyes were intraindividually randomized for treatment with one of the two IOL injectors. Implantation videos were reviewed for inadvertent events and the time for different implantation steps was determined. The injector nozzle tips were examined using light and scanning electron microscopy (SEM). Damage was graded using the Heidelberg Score for IOL injector damage (HeiScore). Three months postoperatively, IOLs were assessed for material changes. Results Implantation was without critical events in 96 of 96 eyes. Mean implantation time was 41.90 ± 7.11 s with the Multisert and 52.22 ± 12.06 s with the AutonoMe. In the AutonoMe group, we observed 4 eyes (8.3%) with a failed docking attempt, 28 eyes (58.3%) with a haptic adherence, one case (2.1%) of straight leading haptic and 2 cases (4.2%) of intrawound IOL manipulation. There were no events observed in the Multisert group. The mean HeiScore values were 0.87 ± 0.61 and 3.68 ± 0.47 for the AutonoMe and Multisert. 3 months postoperatively, IOL material changes were absent. Conclusions Both injectors allowed safe and controlled implantation. Using Multisert, implantation behavior was more consistent. The injectors showed different damage profiles with a higher damage score for the Multisert. The study is registered at the German Clinical Trials Register (Deutsches Register Klinischer Studien; reference number: DRKS00007837).

Funder

Rahel Goitein-Straus-Program of Heidelberg University, Faculty of Medicine

Hoya

Physician-Scientist Program of Heidelberg University, Faculty of Medicine

Klaus Tschira Stiftung

Publisher

SAGE Publications

Subject

Ophthalmology,General Medicine