Diluted contrast media combined with high-resolution C-arm CT for evaluating apposition of flow diversions in a porcine model

Author:

Long Shuhai1ORCID,Yu Sun1,Shi Chengcheng1,Ma Ji1,Li Zhen1,Cheng Jingliang2,Ran Yuncai2,Han Xinwei1,Li Tengfei1ORCID

Affiliation:

1. Department of Interventional Radiology, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China

2. Department of Magnetic Resonance, The First Affiliated Hospital of Zhengzhou University, Zhengzhou, China

Abstract

Background In intracranial stenting, good stent apposition is crucial, and high-resolution C-arm computer tomography (CT) is utilized to assess whether stent apposition is complete. This study was aimed at finding optimal hypertonic or isotonic contrast media injection concentration for high-resolution C-arm CT to assess apposition of flow diversion (FD) after carotid artery stenting in swine. Methods Twelve FD stents were implanted into the left carotid artery of Bama swine through the endovascular method. During high-resolution C-arm CT scanning, 6%, 8%, 10%, 12%, and 14% dilution percentages of hypertonic contrast media (iopromide 370 mg/ml) and 10%, 12%, 14%, 16%, and 18% dilution percentages of isotonic contrast media (iodixanol 320 mg/ml) were separately injected. A radiologist and a neuro-interventional specialist evaluated and qualitatively scored the post-processed images, and intravascular ultrasound (IVUS) was used to verify the accuracy of these images. Results Overall, 12 FD stents were implanted into the left common carotid artery of 12 swine, with a technical success rate of 100%. The best reconstructed images used to observe stent apposition were achieved with iopromide diluted to a concentration of 12% (all P < .05) or iodixanol diluted to a concentration of 16% (all P < .05). Malapposition was noted in one case, and good apposition was noted in 11 cases. These results were consistent with IVUS observations. Conclusion Injecting iopromide or iodixanol diluted to 12% or 16% during high-resolution C-arm CT scanning, respectively, can help evaluate FD stent apposition and obtain optimal image quality.

Funder

Henan Provincial and Ministerial Co-construction Youth Project of Medical Science and Technology Research Plan

Medical Science and Technology Project for Health Commission of Henan Province

National Natural Science Foundation of China

Publisher

SAGE Publications

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3