Effect of the support systems’ compliance on total hip modular taper seating stability

Author:

Scholl Laura1,Pierre David2,Rajaravivarma Raga1,Lee Reginald1,Faizan Ahmad1,Swaminathan Viswanathan1,TenHuisen Kevor1,Gilbert Jeremy L2,Nevelos Jim1

Affiliation:

1. Stryker Orthopaedics, Mahwah, NJ, USA

2. Department of Biomedical and Chemical Engineering, Syracuse University, Syracuse, NY, USA

Abstract

Assembly of a femoral head onto the stem remains non-standardized. The literature shows altering mechanical conditions during seating affects taper strength and lower assembly load may increase fretting corrosion during cyclic tests. This suggests overall performance may be affected by head assembly method. The purpose of this test was to perform bench-top studies to determine influence of peak force magnitude, load rate, and compliance of the system's support structure on initial stability of the taper. Custom manufactured CoCrMo femoral heads and Ti-6Al-4V taper analog samples were assembled with varying peak force magnitudes (2–10.1 kN), load rates (quasi-static vs impaction), and system compliance (rigid vs compliant). A clinically-relevant system compliance design was based off of force data collected during a cadaver impaction study. Tensile loads were then applied to disassemble the taper and quantify initial taper stability. Results indicated that taper stability (assessed by disassembly forces) increased linearly with assembly force and load rate did not have a significant effect on taper stability. When considering system compliance, a 42%–50% larger input energy, dependent on assembly force, was required in the compliant group to achieve a comparable impaction force to the rigid group. Even when this impaction force was achieved, the correlation between the coefficient, defined as distraction force divided by assembly load, was significantly reduced for the compliant test group. The compliant setup was intended to simulate a surgical scenario where patient and surgical factors may influence the resulting compliance. Based on results, surgical procedure and patient variables may have a significant effect on initial taper stability.

Publisher

SAGE Publications

Subject

Mechanical Engineering,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3