Affiliation:
1. Stryker Orthopaedics, Mahwah, NJ, USA
Abstract
Cases of fretting and corrosion at the taper junction have been reported in large metal-on-metal bearing combinations, and more recently, this concern has included metal-on-polyethylene bearing combinations. Many of these patients have been revised due to adverse local tissue reaction secondary to taper corrosion. This taper corrosion–related adverse local tissue reaction seems to be a multifactorial issue and difficult to assess. The aim of this study was to look at one potential variable, the impaction behavior (impaction force, number of blows, etc.) of orthopedic surgeons, and understand how this can affect the locking strength of tapers. A group of experienced orthopedic surgeons were asked to use their typical surgical approach to impact a femoral head onto a hip femoral stem using an Operating Room (OR)-simulated test setup. Impaction parameters such as impaction force, velocity, and energy, as well as the number of impacts, were characterized and applied in a bench-top study used to evaluate the effect of these parameters on the initial stability of the taper junction. High variation was found in the surgical impaction parameters, but overall it was determined that increased impaction force correlated to superior stability of the taper junction.
Subject
Mechanical Engineering,General Medicine
Cited by
30 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献