Influence of a Modified Procedure of Joining Ceramic Head and Adapter Sleeve on the Stem Taper in Revision: An Experimental Study

Author:

Hunger Sandra12ORCID,Rotsch Christian12ORCID,Günther Florian12ORCID,Drossel Welf-Guntram13,Heyde Christoph-Eckhard2

Affiliation:

1. Fraunhofer Institute for Machine Tools and Forming Technology IWU, 01187 Dresden, Germany

2. Department of Orthopaedic, Trauma and Plastic Surgery Clinic, University of Leipzig Medical Center, 04103 Leipzig, Germany

3. Institute for Machine Tools and Production Processes, Faculty of Mechanical Engineering, Chemnitz University of Technology, 09111 Chemnitz, Germany

Abstract

In revision operations, ceramic heads of modular hip implants can be replaced. As the surface of the stem taper can be damaged, additional adapter sleeves are applied. The components are usually connected manually by the surgeon in a one-step procedure by hammer impacts. In this study, we investigated a two-step joining procedure with reproducible impaction force. First, the adapter sleeve and head were joined quasi-statically with a force of 2 kN using an assembly device. In the second step, these components were applied to the stem taper using a pulse-controlled instrument. For reference, the joints were assembled according to standard conditions using a tensile testing machine. An average pull-off force of 1309 ± 201 N was achieved for the components joined by the instrument, and the average measured values for the components joined by the testing machine were 1290 ± 140 N. All specimens achieved a force >350 N when released and therefore met the acceptance criterion defined for this study. This study showed that a modified procedure in two steps with a defined force has a positive effect on the reproducibility of the measured joining forces compared to previous studies.

Funder

Federal Ministry of Education and Research

Publisher

MDPI AG

Reference33 articles.

1. Concepts and clinical aspects of active implants for the treatment of bone fractures;Ganse;Acta Biomater.,2022

2. Wirtz, D.C., Morlock, M., and Schröder, R. (2017). Leitfaden Modulare Revisionsendoprothetik, Springer.

3. Hunger, S., Seidler, A., Rotsch, C., Heyde, C.-E., and Drossel, W.-G. (2023). Evaluating the Feasibility and Reproducibility of a Novel Insertion Method for Modular Acetabular Ceramic Liners. Bioengineering, 10.

4. Merete Medical GmbH (2017). merete BioBall™System: Operationsanleitung und Bestellinformation, Merete Medical GmbH. HDB0001-0114.

5. A580. Minimal invasive revision surgery with modular neck adaptors (BIOBALL);Croce;Orthop. Proc.,2011