Catheter fracture and embolization of the distal fragment are well-known complications of subclavian central venous long-term cannulation. In hemodialysis it is an exceptional event. We report a case of accidental rupture of a cuffed hemodialysis catheter with distal migration of a fragment during a procedure of catheter exchange via guide-wire. According to most reported cases, intravascular catheter separation usually occurs completely asymptomatically; our report confirms that catheter embolization itself is usually asymptomatic. Less than one third of the literature-reported cases have associated symptoms, such as palpitations or chest discomfort. Once diagnosed, treatment is an interventional radiological approach, which has a very high success rate. The replacement of permanent cuffed hemodialysis catheters via guide-wire is a delicate procedure and if catheter embolization is diagnosed, the patient must be referred to a center with specific experience in the retrieval of intravascular objects.

Publisher

SAGE Publications

Subject

Nephrology,Surgery

Cited by 4 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3