Catheter Rupture and Distal Embolisation: A Rare Complication of Central Venous Ports

Author:

Biffi R.1,Pozzi S.1,Orsi F.2,Grasso F.2,De Braud F.3,Cenciarelli S.1,Andreoni B.1

Affiliation:

1. Departments of General Surgery, European Institute of Oncology, Milan - Italy

2. Departments of Radiology, European Institute of Oncology, Milan - Italy

3. Departments of Medical Oncology, European Institute of Oncology, Milan - Italy

Abstract

Central venous access devices placed through a percutaneous subclavian approach may be compressed by neighbouring bony structures, leading to biomaterial fatigue, catheter fracture at the compression site, and possible embolisation of distal fragment into the central veins. The aim of this paper is to review the experience of the authors, including more than 1300 subclavian port placements, carried out during a five-year period, discussing possible causes and therapeutic options of this rare complication. Nine patients out of 1320 (0.68%) experienced this complication during the five-year period of this study. Two patients only showed a retrospective radiologic evidence of the “pinch-off sign” (e.g. initial compression of the catheter at the costo-clavicular junction). No patients had symptoms from the embolised catheter fragment; the most frequent symptom (8 out of 9 cases) was a painful swelling around the port area during infusion, related to the extravasation of medications or fluids into the subcutaneous tissue. The site of embolised segment varied from azygos vein to right pulmonary artery; however, these findings did not affect the outcome, and all the embolised fragments were successfully retrieved through a transfemoral approach using a radiologic interventional technique. No fatality occurred. The catheter fracture and embolisation of the distal fragment are a well-known complication of subclavian central venous long-term cannulation, whose estimated overall incidence is 0.5–1%. Diagnosis is usually based on the radiologic appearance of the catheter compression (so called “pinch-off sign”), which is far from being constant; a clinical suspicion can derive from intermittent malfunction, which claims differential diagnosis with the presence of a fibrin sleeve around the tip of the catheter. Once diagnosed, the treatment is always an interventional radiologic approach, which has a very high success rate. When it fails, the possibility to leave the fragment embolised in the central veins, heart or pulmonary arteries, should be considered, being the thoracotomy and open catheter retraction questionable, at present time, in patients who have no symptoms and limited life-expectancy.

Publisher

SAGE Publications

Subject

Nephrology,Surgery

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3