Hemodialysis vascular access flow measurements by the novel DMed NephroFlow® device: A comparative study with Transonic®

Author:

Merino Jose L1ORCID,García Esther1,Varillas-Delgado David2,Mendoza Sonia1,Bueno Blanca1,Domínguez Patricia1,Bucalo Laura1,Espejo Beatriz1,Baena Laura1,Paraíso Vicente1

Affiliation:

1. Sección Nefrología, Hospital Univ, del Henares, Coslada, Madrid, Spain

2. Facultad de Ciencias de la Salud, Universidad Francisco de Vitoria, Madrid, Spain

Abstract

Introduction: The current Spanish Clinical Guidelines on Vascular Access for Hemodialysis support the need for surveillance and monitoring of vascular access (VA) to avoid complications. Ultrasound dilution (UD) methods are accepted for the evaluation of VA flow and Transonic® has established the gold standard method for the measurement. The DMed NephroFlow (NIPRO®) device, based on UD method has recently been incorporated. We report a comparative study between the classic Transonic® versus the new NephroFlow® device. Material and methods: For two consecutive months, measurements of VA flow using both referred systems were performed in patients with a native arteriovenous fistula (AVF) or a graft (AVG) on hemodialysis (HD) in our unit. Both studies were undertaken according to the usual recommendations: VA flow of 250 ml/min, ultrafiltration rate without modifications, both needles in the same vein, and always in the first hour of the HD session. Results: Forty-five patients were included: 17 women and 28 men, mean age of 67 ± 12 years. Thirty patients were diabetic. The baseline meantime on HD was 51 ± 39 months (range: 3–163). Type of VA was: 17 patients radio-cephalic AVF, 17 brachiocephalic AVF, 7 brachiobasilic AVF, and 3 with a graft. The mean flow estimated by the Transonic® was 1222 ± 805 ml/min and the estimated flow by the NephroFlow® device was 1252 ± 975 ml/min. Good reliability between Transonic® and NephroFlow® was observed, with a reliability index of Cronbach’s Alpha of 0.927 and an Intraclass Correlation Index of 0.928. Conclusions: The NephroFlow® device seems comparable with the accepted gold standard UD method for estimating VA flow. More studies must be performed to verify these results. However, they should be considered for the surveillance and monitoring of VA flow, in agreement with the Spanish Guidelines.

Publisher

SAGE Publications

Subject

Nephrology,Surgery

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3