Affiliation:
1. Department of Electrical and Computer Engineering, Hajim School of Engineering and Applied Sciences, University of Rochester, Rochester, NY, USA
Abstract
Imaging plaque microvasculature with contrast-enhanced intravascular ultrasound (IVUS) could help clinicians evaluate atherosclerosis and guide therapeutic interventions. In this study, we evaluated the performance of chirp-coded ultraharmonic imaging using a modified IVUS system (iLab™, Boston Scientific/Scimed) equipped with clinically available peripheral and coronary imaging catheters. Flow phantoms perfused with a phospholipid-encapsulated contrast agent were visualized using ultraharmonic imaging at 12 MHz and 30 MHz transmit frequencies. Flow channels with diameters as small as 0.8 mm and 0.5 mm were visualized using the peripheral and coronary imaging catheters. Radio-frequency signals were acquired at standard IVUS rotation speed, which resulted in a frame rate of 30 frames/s. Contrast-to-tissue ratios up to 17.9 ± 1.11 dB and 10.7 ± 2.85 dB were attained by chirp-coded ultraharmonic imaging at 12 MHz and 30 MHz transmit frequencies, respectively. These results demonstrate the feasibility of performing ultraharmonic imaging at standard frame rates with clinically available IVUS catheters using chirp-coded excitation.
Subject
Radiology Nuclear Medicine and imaging,Radiological and Ultrasound Technology
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献