Manually Steerable Catheter With Improved Agility

Author:

Tangen Geir Arne123,Manstad-Hulaas Frode124,Nypan Erik12,Brekken Reidar123

Affiliation:

1. Department of Circulation and Medical Imaging, Norwegian University of Science and Technology (NTNU), Trondheim, Norway

2. Norwegian National Advisory Unit for Ultrasound and Image-Guided Therapy, St. Olavs Hospital, Trondheim, Norway

3. Department of Health Research, SINTEF Technology and Society, Trondheim, Norway

4. Department of Radiology, St. Olavs Hospital, Trondheim, Norway

Abstract

Purpose: A prototype steerable catheter was designed for endovascular procedures. This technical pilot study reports the initial experience using the catheter for cannulation of visceral arteries. Technique: The 7F catheter was manually steerable with operator control handle for bending and rotation of the tip. The maximum bending angle was approximately 90° and full 360° rotation of the tip was supported. The study involved 1 pig with 4 designated target arteries: the left and right renal arteries, the superior mesenteric artery, and the celiac trunk. Fluoroscopy with 3-dimensional (3D) overlay showing the ostia from preoperative computed tomography angiography was used for image guidance. The cannulation was considered successful if the guidewire was placed well inside the target artery. In addition to evaluating cannulation success, procedure time and associated radiation doses were recorded. The procedure was performed twice with 2 different operators. Conclusions: Both operators successfully reached all 4 target arteries, demonstrating the feasibility of the steerable catheter for endovascular cannulation of visceral arteries. No contrast medium was used, and median radiation dose was 4.5 mGy per cannulation. An average of approximately 2 minutes was used per cannulation. This study motivates further testing in a more comprehensive study to evaluate reproducibility in several animals and with inclusion of more operators. Further development by integrating the new catheter tool in a navigation system is also an interesting next step, combining fine control of catheter tip movements and 3D image guidance without ionizing radiation.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine

Cited by 5 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3