Failure Mode of Suture Anchors as a Function of Insertion Depth

Author:

Bynum C. Kelly1,Lee Steven2,Mahar Andrew23,Tasto James2,Pedowitz Robert2

Affiliation:

1. Rocky Mountain Orthopaedic Associates, Grand Junction, Colorado

2. San Diego Arthroscopy and Sports Medicine Fellowship at University of California San Diego, San Diego, California

3. Orthopedic Biomechanics Research Center, Children's Hospital, San Diego, California

Abstract

Background Surgeons can control not only the angle but also the depth of suture anchor placement during arthroscopic rotator cuff repair, although the tendency may be to place suture anchors on the deep side to avoid damage from prominent anchor eyelets. However, little information is available regarding possible effects of suture anchor depth on construct failure mechanisms. Hypothesis Anchor depth affects the mode of suture failure with physiologically relevant cyclic loads. Study Design Controlled laboratory study. Methods Metallic screw-in suture anchors loaded with No. 2 braided polyester sutures were inserted into the bovine infra-spinatus footprint with the eyelet proud, standard, or deep. Sutures were hand tied to create a closed loop. Constructs were cyclically loaded from 10 to 90 N and, if still intact at 500 cycles, taken to ultimate failure (maximum load). Results When clinical failure was defined as greater than 3-mm construct elongation, anchors placed with the eyelet deep experienced statistically earlier clinical failure via cutting of the suture through the bone (P <. 02). However, anchors placed at this level did not experience catastrophic failure during cyclic loading. The standard and proud anchors experienced 3 mm of elongation at a greater number of cycles, but the suture material degraded at the anchor eyelet, and a majority of these constructs broke during cyclic physiologic loading. At failure testing, the deep anchors had a significantly increased failure load (164 N) compared to standard (133 N) (P <. 04) and proud (113 N) anchors (P <. 005). Conclusion Varying the depth of suture anchor insertion changes the mechanical properties and mode of failure of suture anchor constructs. Clinical Relevance Surgeons should be aware of the effects of suture anchor depth and abrasive eyelet wear on construct failure during arthroscopic rotator cuff repair.

Publisher

SAGE Publications

Subject

Physical Therapy, Sports Therapy and Rehabilitation,Orthopedics and Sports Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3