Affiliation:
1. New Orleans, Louisiana
Abstract
The ignition of an endotracheal cuff by CO2 laser impaction has proved troublesome in several ignition sequences. Metal foil tape wrapped around the shaft of the tube will provide adequate protection for it, but the tape does not extend to the thin-walled cuff. When normal saline solution is used instead of air to fill the cuff, heat is rapidly conducted away from the wall and the cuff is prevented from reaching its kindling temperature. If penetration should occur, a fine jet of saline solution instantaneously sprays the field, extinguishing any early combustion. The reaction time of the surgeon or an assistant is not a factor in prevention of thermal injury, no new equipment is required and preparation time is minimal.
Subject
General Medicine,Otorhinolaryngology
Cited by
44 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献