Abstract
The ignition characteristics of the American Hospital, Portex®, polyvinyl chloride, and red rubber tracheal tubes commonly available have been evaluated using a standard carbon dioxide laser. The added hazard of the increased combustibility of the impregnated line on endotracheal tubes and the hazard of commonly utilized anesthetic or lubricating materials are discussed. The safety limits which, experimentally, indicate levels of energy and laser burst time which seem to preclude the hazard of ignition are outlined. The potential danger of postburn bronchiolitis is noted, along with appropriate treatment.
Subject
General Medicine,Otorhinolaryngology
Cited by
20 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献