Abstract
The use of Gore-Tex (expanded polytetrafluoroethylene) for medialization laryngoplasty has recently been described. Clinically, this laryngeal implant has performed well, but no animal studies evaluating the use of this implant in the larynx have been published. This study examines the gross and histologic response of the rabbit larynx to the Gore-Tex implant. Five rabbits underwent bilateral medialization with the Gore Thyroplasty Device. The vocal folds were left innervated and mobile. After 6 months, the rabbits were painlessly sacrificed and the implants were evaluated by gross inspection and histologic examination. No gross changes to the glottis occurred, and the implants were easily removed with gentle traction. A modest inflammatory response with a thin fibrous capsule was present around the implant. Foreign body giant cells were observed at the interface of the implant with the tissue. No migration or extrusion of the implant was observed. These results suggest that Gore-Tex is biocompatible in the rabbit larynx.
Subject
General Medicine,Otorhinolaryngology
Cited by
23 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献