Abstract

Implantable prosthetic ventricles and trileaflet valves made by vacuum forming have been developed and implant tested. All components are made from Pellethane®. Recognizing the need for smaller as well as larger ventricles, designs with effective stroke volumes of 50, 85, 100 and 130 cc have been tested with several valve types. The pneumatically driven Utah ventricular assist device (UVAD) can be used as a total artificial heart (TAH) or ventricular assist device (VAD) by using the appropriate inflow and outflow adapters. In vitro durability testing has demonstrated ventricular lifetime beyond two years and valve lifetime to nearly one and one half years. The polymer valves have lower regurgitation than mechanical valves. Animal implantation experience includes 21 TAH implants and 16 left ventricular assist device (LVAD) implantations. TAH survival ranges from 2 to 210 days. LVAD animals have lived up to 116 days before elective termination. The animal were healthy and grew normally. The devices exhibit a “Starling's Law” response. One TAH animal survived 72 days before successful explantation followed by transplantation. At autopsy, this animal had no renal infarcts. Hematology data has demonstrated the existence of little or no intravascular hemolysis (PF Hb < 5 mg%). The “Philadelphia” version of the UVAD vacuum formed ventricles are small enough to be implanted without thrombus provoking connectors. Eight animals have received this TAH and survived up to 120 days. Vacuum forming offers a rapid and inexpensive way to produce reliable and effective total artificial hearts and valves for widespread, temporary clinical application in any size adult human.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Reference12 articles.

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3