Patient-specific design process and evaluation of a hip prosthesis femoral stem

Author:

Abdelaal Osama12,Darwish Saied3,El-Hofy Hassan4,Saito Yoshio5

Affiliation:

1. Department of Mechanical and Industrial Engineering, College of Engineering, Majmaah University, Majmaah, Saudi Arabia

2. Department of Mechanical Engineering, Faculty of Engineering, Assiut University, Assiut, Egypt

3. Ministry of Higher Education & Scientific Research, Cairo, Egypt

4. Industrial Engineering and Systems Management Department (IESM), School of Innovative Design Engineering (IDE), Egypt-Japan University for Science and Technology (E-JUST), Alexandria, Egypt

5. Department of Mechanical and Control Engineering, Tokyo Institute of Technology, Tokyo, Japan

Abstract

Introduction: There are several commercially available hip implant systems. However, for some cases, custom implant designed based on patient-specific anatomy can offer the patient the best available implant solution. Currently, there is a growing trend toward personalization of medical implants involving additive manufacturing into orthopedic medical implants’ manufacturing. Methods: This article introduces a systematic design methodology of femoral stem prosthesis based on patient’s computer tomography data. Finite element analysis is used to evaluate and compare the micromotion and stress distribution of the customized femoral component and a conventional stem. Results: The proposed customized femoral stem achieved close geometrical fit and fill between femoral canal and stem surfaces. The customized stem demonstrated lower micromotion (peak: 21 μm) than conventional stem (peak: 34 μm). Stress results indicate up to 89% increase in load transfer by conventional stem than custom stem because the higher stiffness of patient-specific femoral stem proximally increases the custom stem shielding in Gruen’s zone 7. Moreover, patient-specific femoral stem transfers the load widely in metaphyseal region. Conclusion: The customized femoral stem presented satisfactory results related to primary stability, but compromising proximo-medial load transfer due to increased stem cross-sectional area increased stem stiffness.

Publisher

SAGE Publications

Subject

Biomedical Engineering,Biomaterials,General Medicine,Medicine (miscellaneous),Bioengineering

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Method of computational design for additive manufacturing of hip endoprosthesis based on basic‐cell concept;International Journal for Numerical Methods in Biomedical Engineering;2024-01-21

2. Structural Behavior Analysis of the Proximal Femur after Artificial Hip Joint Implantation at Micro Level;Journal of the Korean Society for Precision Engineering;2023-11-01

3. Bimodular femoral stems in primary total hip arthroplasty;Expert Review of Medical Devices;2023-10-09

4. IMPACT OF GEOMETRIC PARAMETERS ON THE CHOICE OF A CUSTOM FEMORAL IMPLANT;Journal of Mechanics in Medicine and Biology;2023-07-29

5. Design of titanium uncemented femoral stems for hip prosthesis suitable for the Colombian young adult population;Computer Methods in Biomechanics and Biomedical Engineering;2023-05-05