Affiliation:
1. Biomechanics Laboratory, Division of Orthopedic Research, Mayo Clinic, Rochester, Minnesota, USA
2. Department of Orthopedic Surgery, Universitair Ziekenhuis Antwerpen, Antwerp, Belgium
Abstract
Purpose. To evaluate the effect of prosthesis design on stress profile in the proximal femur after hip resurfacing. Methods. The von Mises stress profile of the native femur was simulated and compared with that of resurfaced femurs using various prosthetic materials (titanium, cobalt-chrome, ceramic), stem lengths (normal, half, short, and no stem), and femoral head coverage (shell size) [260°, 220°, 180°, and 140°]. Results. Hip resurfacing altered the stress profile of the cancellous (but not cortical) bone of the femoral neck. Maximal cortical stresses were observed at the posterior half of the medial femoral neck. The stress profile of the native femur was most similar to that of the resurfaced femur made of titanium, with a short or no stem and 260 degrees of femoral head coverage (shell size). Conclusion. Optimising prosthesis design by minimising biomechanical alterations seems a valid approach to achieving favourable long-term outcomes. Cadaveric and in vivo studies are needed to confirm the clinical relevance and feasibility.
Cited by
11 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献