Improvement in Proximal Aortic Endograft Fixation: An Experimental Study Using Different Stent-Grafts in Human Cadaveric Aortas

Author:

Veerapen Reuben,Dorandeu Anne1,Serre Isabelle2,Berthet Jean-Phillipe,Marty-Ane Charles Henri,Mary Henri,Alric Pierre

Affiliation:

1. Department of Legal Medicine, Hôpital Arnaud de Villeneuve, CHU Montpellier, France

2. Department of Pathology, Hôpital Arnaud de Villeneuve, CHU Montpellier, France

Abstract

Purpose: To compare the proximal fixation characteristics of different types of stent-grafts (SG) and the adjunctive effect of a proximally placed Palmaz stent. Methods: Human cadaveric aortas were obtained at autopsy and cut into 30 aortic segments. Appropriately oversized (10% to 20%) commercial stent-grafts (Zenith, Ancure, Excluder, Talent) and 4 configurations of homemade Palmaz-based stent-grafts (polytetrafluoroethylene or polyester coverings each with no uncovered proximal edge or a 20-mm bare section of stent) were implanted 20 mm into an aortic segment and balloon dilated. Each segment was placed in an experimental apparatus to measure the dislodgment force. Bare Palmaz stents were used to reinforce the proximal fixation of the commercial stent-grafts, and the measurements were repeated. Manual anastomoses were made and their dislodgment force tested for comparison. Results: The median dislodgment force ranged from 6.5 N for the Excluder to 26.5 N for the Zenith (8.0 N for the Talent, 11.8 N for the Ancure, and 8.1 to 10.7 N for the various Palmaz stent-graft designs). There was no significant difference between the Zenith and the Ancure groups or between the Excluder and the Talent groups. However, the Zenith and Ancure devices required significantly higher (p = 0.0004) force (∼25%) to displace them than the other stent-grafts tested. With the Palmaz stent added to the proximal attachment site, the median dislodgment force for the different commercial endografts was significantly improved for all devices (p < 0.03): 34% for the Zenith device, 69% for the Ancure, 73% for the Talent, and 80% for the Excluder endoprosthesis. Conclusions: The Zenith stent-graft had the best resistance to dislodgment. An additional Palmaz stent placed at the proximal attachment site greatly improves endograft fixation regardless of the type of stent-graft. For complex aneurysm necks or for intraoperative type I endoleak management, an adjunctive Palmaz stent could be used.

Publisher

SAGE Publications

Subject

Cardiology and Cardiovascular Medicine,Radiology Nuclear Medicine and imaging,Surgery

Cited by 19 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3