Affiliation:
1. Biotherapeutics Discovery, Boehringer Ingelheim Pharmaceuticals, 900 Ridgebury Road, Ridgefield, CT 06877, USA
Abstract
Aim: The fully automated microfluidics-based Gyrolab is a popular instrument for the bioanalysis of protein therapeutics; requiring minimal sample and reagent volumes. Gyros offers affinity software for determining binding affinity in solution using a high-throughput method and miniaturized reactions. Results: Using this affinity software, multiple CTGF-targeting reagents were characterized on the Gyrolab after <100% target coverage was seen in a cynomolgus pharmacokinetic/PD study dosed with anti-CTGF antibodies. The results uncovered magnitude differences in binding affinities between the dosed antibody, target and assay reagents. Conclusion: The binding affinity values were used to investigate reduced target coverage and results highlight potential of the affinity software for incorporation into the bioanalyst's existing Gyrolab workflow for characterizing reagents and optimizing pharmacokinetic/PD bioanalytical assays.
Subject
Medical Laboratory Technology,Clinical Biochemistry,General Pharmacology, Toxicology and Pharmaceutics,General Medicine,Analytical Chemistry
Cited by
3 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献