Affiliation:
1. Department of Engineering Mechanics, School of Aerospace, Tsinghua University, Beijing 100084, P. R. China
Abstract
Low hemolysis is an important factor for axial blood pumps that has been used in patients with heart failure. The structure of impellers plays a key role in the hemolytic properties of axial blood pumps. Axial blood pumps with various structure impellers exhibit different hemolytic characteristic. In the present study, we aimed to investigate the type of impellers structures in axial blood pumps that contain the best low hemolytic properties. Also, it is expensive and time-consuming to validate the axial blood pump's hemolytic property by in vivo experiments. Therefore, in the present study, the numerical method was applied to analyze the hemolytic property in a blood pump. Specifically, the hemolysis of the pump was calculated by using a forward Euler approach based on the changes in shear stress and related exposure times along the particle trace lines. The different vane structures and rotational speed that affect hemolysis were analyzed and compared. The results showed that long–short alternant vanes exhibited the best hemolytic property which could be utilized in the optimization design of axial blood pumps.
Publisher
World Scientific Pub Co Pte Lt
Cited by
9 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献