Abstract
AbstractAdvances in single-particle cryogenic electron microscopy (cryoEM) now allow for routine structure determination of well-behaved biological specimens to high-resolution. Despite advances in the electron microscope, direct electron detectors, and data processing software, the preparation of high-quality grids with thin layers of vitreous ice containing the specimen of interest in random orientations remains a critical bottleneck for many projects. Although numerous efforts have been dedicated to overcoming hurdles frequently encountered during specimen vitrification using traditional blot-and-plunge specimen preparation techniques, the development of blot-free grid preparation devices provide a unique opportunity to carefully tune ice thickness, particle density, and specimen behavior during the vitrification process for improvements in image quality. Here, we describe critical steps of high-quality grid preparation using a SPT Labtech chameleon, evaluation of grid quality/ice thickness using the chameleon software, high-throughput imaging in the electron microscope, and recommend steps for troubleshooting grid preparation when standard parameters fail to yield suitable specimen.Video LinkContents of this manuscript are available as a video tutorial. This video can be foundhere