Author:
Miyai K,Ishibashi K,Kawashima M
Abstract
Abstract
We describe a sensitive, simple, and rapid two-site immunoenzymometric assay for thyrotropin in dried blood samples on filter paper, for use in screening for neonatal primary hypothyroidism. In this method, two dried-blood spots of 3 mm diameter (equivalent to about 5.4 microliter of blood) are incubated overnight with anti-thyrotropin-beta-D-galactosidase complex in an anti-thyrotropin coated tube. Then the enzyme activity in the washed tube is determined fluorophotometrically. The range of thyrotropin measurable is 10 to 160 milli-int. units/L blood. Values for thyrotropin in dried blood samples determined by this method and those determined by radioimmunoassay correlated highly (r = 0.96).
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Cited by
12 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献