Affiliation:
1. Department of Laboratory Medicine, Box 359743, Harborview Medical Center and University of Washington, Seattle, WA 98104-2499
Abstract
Abstract
Given the widespread interest in antioxidant nutrients, we have developed a new procedure that will permit the automated determination of plasma ascorbic acid (AA) concentration with a Roche Fara centrifugal analyzer. After the deproteinization of plasma with metaphosphoric acid, AA is oxidized to dehydroascorbic acid by AA oxidase. The product is coupled to o-phenylenediamine to produce a chromophore for which the absorbance is measured at 340 nm. The procedure allows much faster throughput than conventional HPLC methods while yielding results that correlate well and provide improved precision. Linearity extends beyond the reference range of 26.1–84.6 μmol/L, and severe hemolysis is the only interference identified.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry