Candidate Reference Method for Total Thyroxine in Human Serum

Author:

Tai Susan S-C1,Sniegoski Lorna T1,Welch Michael J

Affiliation:

1. Analytical Chemistry Division, National Institute of Standards and Technology, Gaithersburg, MD 20899-8392

Abstract

Abstract Background: There is a need for a critically evaluated reference method for thyroxine to provide an accuracy base to which routine methods can be traceable. We describe a candidate reference method involving isotope-dilution coupled with liquid chromatography/mass spectrometry. Methods: An isotopically labeled internal standard, thyroxine-d5, was added to serum, followed by equilibration, protein precipitation, and ethyl acetate and solid-phase extractions to prepare samples for liquid chromatography-mass spectrometry electrospray ionization (LC/MS-ESI) analysis. For separation, a Zorbax Eclipse XDB-C18 column was used with a mobile phase consisting of 1 mL/L acetic acid in acetonitrile-water (32:68 by volume) for positive ions and a Zorbax Extend-C18 column with a mobile phase consisting of 2 mL/L ammonium hydroxide in methanol-water (32:68 by volume) for negative ions. [M + H]+ ions at m/z 778 and 783 for thyroxine and its labeled internal standard were monitored for positive ions and [M − H]− ions at m/z 776 and 781 for negative ions. Samples of frozen serum pools were prepared and measured in three separate sets. Results: Within-set CVs were 0.2–1.0%. The correlation coefficients of all linear regression lines (measured intensity ratios vs mass ratios) were 0.999–1.000. Positive- and negative-ion measurements agreed with a mean difference of 0.45% at three concentrations (50, 110, and 168 μg/L). The detection limits (at a signal-to-noise ratio of ∼3 to 5) were 30 and 20 pg for positive and negative ions, respectively. The results from the LC/MS-ESI method were within 1 SD of the composite means from many routine clinical methods, although it appears that the clinical method means may be biased high by 4–5 μg/L across the concentrations. Some routine clinical methods may be biased by up to 20% at low concentrations. Conclusions: This well-characterized LC/MS-ESI method for total serum thyroxine with a theoretically sound approach, demonstrated good accuracy and precision, and low susceptibility to interferences qualifies as a candidate reference method. Use of this reference method as an accuracy base may reduce the apparent biases in routine methods along with the high interlaboratory imprecision.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3