Author:
DeRuyter M G,De Leenheer A P
Abstract
Abstract
We propose a single-run liquid-chromatographic determination, with ultraviolet detection at 330 nm, for serum retinol and retinyl esters. The vitamin A derivatives are extracted according to the Bligh-Dyer procedure. With 200 microliter or serum, the lower detection limit is 50 microgram/liter for retinol and about 100 microgram/liter for retinyl esters. Within-run precision (CV) was 2.3% for retinol, 4.3% for retinyl palmitate. Day-to-day percision (CV, n = 20) for retinol was 4.9% during a month. The method can be used for the assessment of vitamin A absorption tests and for the determination of serum retinol (normal, subnormal, and above-normal concentrations). Serum retinyl esters can only be measured in conditions where concentrations exceed 100 microgram/liter.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
78 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献