Impact of electroanatomical mapping-guided lead implantation on procedural outcome of His bundle pacing

Author:

Richter Sergio1,Ebert Micaela1,Bertagnolli Livio1,Gebauer Roman2,Lucas Johannes1,Scheller Dominik3,Paetsch Ingo1,Hindricks Gerhard1,Döring Michael1

Affiliation:

1. Department of Electrophysiology, Heart Center, University of Leipzig, Strümpellstr. 39, 04289 Leipzig, Germany

2. Department of Pediatric Cardiology, Heart Center, University of Leipzig, Leipzig, Germany

3. Department of Therapy Specialists, Electrophysiology, Abbott Medical GmbH, Eschborn, Germany

Abstract

Abstract Aims Conventional His bundle pacing (HBP) can be technically challenging and fluoroscopy-intense, particularly in patients with His-Purkinje conduction disease (HPCD). Three-dimensional electroanatomical mapping (EAM) facilitates non-fluoroscopic lead navigation and HB electrogram mapping. We sought to assess the procedural outcome of routine EAM-guided HBP compared with conventional HBP in a real-world population and evaluate the feasibility and safety of EAM-guided HBP in patients with HPCD. Methods and results  We included 58 consecutive patients (72 ± 13 years; 71% male) who underwent an attempt to conventional (EAM− group; n = 29) or EAM-guided (EAM+ group; n = 29) HBP between June 2019 and April 2020. The centre’s learning curve was initially determined (n = 40 cases) to define the conventional control group and minimize outcome bias favouring EAM-guided HBP. His bundle pacing was successful in 26 patients (90%) in the EAM+ and 27 patients (93%) in the EAM− group (P = 0.64). The procedure time was 90 (73–135) and 110 (70–130) min, respectively (P = 0.89). The total fluoroscopy time [0.7 (0.5–1.4) vs. 3.3 (1.4–6.5) min; P < 0.001] and fluoroscopy dose [21.9 (9.1–47.7) vs. 78.6 (27.2–144.9) cGycm2; P = 0.001] were significantly lower in the EAM+ than EAM− group. There were no significant differences between groups in His capture threshold (1.2 ± 0.6 vs. 1.4 ± 1.0 V/1.0 ms; P = 0.33) and paced QRS duration (113 ± 15 vs. 113 ± 17 ms; P = 0.89). In patients with HPCD, paced QRS duration was similar in both groups (121 ± 15 vs. 123 ± 12 ms; P = 0.77). The bundle branch-block recruitment threshold tended to be lower in the EAM+ than EAM− group (1.3 ± 0.7 vs. 1.8 ± 1.2 V/1.0 ms; P = 0.31). No immediate procedure-related complications occurred. One patient (2%) experienced lead dislodgement during 4-week follow-up. Conclusion  Implementation of routine EAM-guided HBP lead implantation is feasible and safe in a real-world cohort of patients with and without HPCD and results in a tremendous reduction in radiation exposure without prolonging procedure time or increasing procedure-related complications.

Publisher

Oxford University Press (OUP)

Subject

Physiology (medical),Cardiology and Cardiovascular Medicine

Reference20 articles.

1. 2018 ACC/AHA/HRS guideline on the evaluation and management of patients with bradycardia and cardiac conduction delay: executive summary: a report of the American College of Cardiology/American Heart Association Task Force on Clinical Practice Guidelines, and the Heart Rhythm Society;Kusumoto;Circulation,2019

2. The effects of right ventricular apical pacing on ventricular function and dyssynchrony: implications for therapy;Tops;J Am Coll Cardiol,2009

3. Permanent his-bundle pacing is feasible, safe, and superior to right ventricular pacing in routine clinical practice;Sharma;Heart Rhythm,2015

4. Clinical outcomes of his bundle pacing compared to right ventricular pacing;Abdelrahman;J Am Coll Cardiol,2018

5. Permanent his-bundle pacing as an alternative to biventricular pacing for cardiac resynchronization therapy: a multicenter experience;Sharma;Heart Rhythm,2018