Evaluation of Quadrox-i® Adult Hollow Fiber Oxygenator with Integrated Arterial Filter

Author:

Guan Yulong,Su Xiaowei,McCoach Robert,Wise Robert,Kunselman Allen,Ündar Akif

Abstract

Gaseous microemboli (GME) remain a challenge for cardiopulmonary bypass procedures in adult as well as pediatric cardiac surgery patients. The present study tested the effectiveness of a new adult membrane oxygenator in models both with and without an integrated arterial filter to evaluate GME trapping capability and determine membrane pressure drops at various flow rates and temperatures. The experimental circuit included a RotaFlow centrifugal blood pump, Quadrox-i® (n= 8) or Quadrox® (n= 8) adult microporous membrane oxygenator, and Sorin adult tubing package. A Sorin Cardiovascular® VVR® 4000i venous reservoir served as pseudo-patient. The circuit was primed with 900 mL heparinized human red blood cells and 300 mL Lactated Ringer’s solution. The final hematocrit was 36%. Tests were performed at different flow rates (4 L/min, 5 L/min, and 6 L/min) and temperatures (35° and 30°). Five mL of bolus air was injected into the venous line over 15 seconds using a syringe connected to a 3/8 × 1/2 luer connector. The Quadrox-i® adult microporous membrane oxygenator with integrated arterial filter had a similar pressure drop at 4 L/min and 35°C compared with Quadrox® membrane oxygenator whereas it had higher pressure drops at 5 L/min and 6 L/min (p<.001). Quadrox-i® adult microporous membrane oxygenator reduced the total emboli count and total emboli volume delivered to the pseudo-patient at all flow rates (p<.001). The emboli handling of Quadrox-i® adult microporous membrane oxygenator was not affected by flow rate and temperature. Compared with the traditional Quadrox® oxygenator, Quadrox-i® adult microporous membrane oxygenator with integrated arterial filter and Softline coating has improved GME handling capacity.

Publisher

EDP Sciences

Subject

Cardiology and Cardiovascular Medicine,Health Professions (miscellaneous),Medicine (miscellaneous)

Cited by 3 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Evaluation of CPB devices relative to their capabilities of reducing the transmission of gaseous microemboli (GME) to a patient during cardiopulmonary bypass;ANSI/AAMI/ISO TIR19024:2016; Evaluation of CPB devices relative to their capabilities of reducing the transmission of gaseous microemboli (GME) to a patient during cardiopulmonary bypass;2016-09-02

2. Effects of Purge-Flow Rate on Microbubble Capture in Radial Arterial-Line Filters;The Journal of ExtraCorporeal Technology;2016-09

3. The Effects of Pressure on Gases in Solution: Possible Insights to Improve Microbubble Filtration for Extracorporeal Circulation;The Journal of ExtraCorporeal Technology;2013-06

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3