The dialysis circuit was washed by the ultrafine bubble (UFB) water and ozone UFB water and sodium hypochlorite aqueous solution. The effect of washing time, ozone concentration of UFB water, influence of UFB water on artificial dialysis equipment, etc. were investigated. The cleaning and sterilizing effect of ozone UFB water passed through the filter for artificial dialysis was verified. It was confirmed that there were no colony bacteria in the circuit which was washed by the ozone UFB water for 1 hour. Ozone UFB water has the same bactericidal effect as sodium hypochlorite. As an alternative of sodium hypochlorite, we are expecting its application to the cleaning and sterilization
of ozone UFB water for dialysis circuit.