Abstract
In order to test the response of radiation-induced current with wide range of dose rate, a Silicon Carbide (SiC) dosimeter is exposed to gamma-rays emitted from a60Co source. The SiC dosimeter in this study is made of a high purity semi-insulating 4H-SiC with nickel and aluminum electrodes. We have successfully demonstrated that the radiation-induced currents in the dosimeter show a linear relationship with the dose rate, and are repeatable and stable.
Publisher
Trans Tech Publications, Ltd.
Subject
Mechanical Engineering,Mechanics of Materials,Condensed Matter Physics,General Materials Science
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献