Author:
Schafer Erin C.,Thibodeau Linda M.
Abstract
Speech recognition was evaluated for ten adults with normal hearing and eight adults with Nucleus cochlear implants (CIs) at several different signal-to-noise ratios (SNRs) and with three frequency modulated (FM) system arrangements: desktop, body worn, and miniature direct connect. Participants were asked to repeat Hearing in Noise Test (HINT) sentences presented with speech noise in a classroom setting and percent correct word repetition was determined. Performance was evaluated for both normal-hearing and CI participants with the desktop soundfield system. In addition, speech recognition for the CI participants was evaluated using two FM systems electrically coupled to their speech processors. When comparing the desktop sound field and the No-FM condition, only the listeners with normal hearing made significant improvements in speech recognition in noise. When comparing the performance across the three FM conditions for the CI listeners, the two electrically coupled FM systems resulted in significantly greater improvements in speech recognition in noise relative to the desktop soundfield system.