Spinal duraplasty materials and hydrostasis: a biomechanical study

Author:

Finn Michael A.1,Faulkner Nathan D.2,Hetzel Scott J.3,Anderson Paul A.2

Affiliation:

1. 1Department of Neurosurgery, University of Colorado School of Medicine, Aurora, Colorado; and

2. 2Departments of Orthopedic Surgery and Rehabilitation Medicine and

3. 3Biostatistics and Medical Informatics, University of Wisconsin, Madison, Wisconsin

Abstract

Object Dural tears are a frequent complication of spinal surgery and contribute to significant morbidity. Occasionally, dural tears cannot be closed primarily and dural patch grafts must be utilized. No data exist on the comparative immediate hydrostatic strength of various patch materials used alone or with a biological adhesive in a spinal dural tear model. Thus, the authors conducted this study to determine the comparative effectiveness of various patch materials used with and without biological adhesive. Methods Twenty-four thoracic spines from calves were prepared with laminectomies and spinal cord evacuation, leaving the dura intact. Foley catheters were inflated on either side of a planned dural defect, and baseline hydrostasis was measured using a fluid column at 30, 60, and 90 cm of H2O. A standard dural defect (1 × 2 cm) was created, and 8 patches of each material (human fascia lata, Duragen, and Preclude) were sutured in place using 5-0 Prolene hemo-seal running sutures. Hydrostasis was again tested at the same pressures. Finally, a hydrogel sealant (Duraseal) was placed over the defect and hydrostasis was again tested. Results were analyzed with repeated measures ANOVA. Results The leakage rate increased significantly at each pressure tested for all conditions. There was no difference in leakage among the 3 patch materials at any of the pressures or for either condition (with or without sealant). All patch materials allowed significantly greater leakage than the intact condition at all pressures. The use of sealant reduced leakage significantly at the 30 and 60 mm Hg pressures to levels similar to the intact condition. At 90 mm Hg, leakage of the sealed construct was greater than at the intact condition but significantly less than without the use of sealant. Conclusions All 3 dural patch materials were of similar hydrostatic strength and allowed greater leakage than at the intact condition. The use of sealant reduced the amount of leakage at all pressures compared with patching alone but allowed more leakage than the intact state at a high hydrostatic pressure (90 mm Hg).

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 8 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3