Neuroendovascular-specific engineering modifications to the CorPath GRX Robotic System

Author:

Britz Gavin W.1,Panesar Sandip S.1,Falb Peter2,Tomas Johnny2,Desai Virendra1,Lumsden Alan3

Affiliation:

1. Department of Neurological Surgery and Neurological Institute, and

2. Corindus, Inc., Waltham, Massachusetts

3. Department of Cardiovascular Surgery, Houston Methodist Hospital, Texas Medical Center, Houston, Texas; and

Abstract

OBJECTIVEThe aim of this study was to evaluate new, neuroendovascular-specific engineering and software modifications to the CorPath GRX Robotic System for their ability to support safer and more effective cranial neurovascular interventions in a preclinical model.METHODSActive device fixation (ADF) control software, permitting automated manipulation of the guidewire relative to the microcatheter, and a modified drive cassette suitable for neuroendovascular instruments were the respective software and hardware modifications to the current CorPath GRX robot, which was cleared by the FDA for percutaneous coronary and peripheral vascular intervention. The authors then trialed the modified system in a live porcine model with simulated neuroendovascular pathology. Femoral access through the aortic arch to the common carotid artery was accomplished manually (without robotic assistance), and the remaining endovascular procedures were performed with robotic assistance. The system was tested for the enhanced ability to navigate and manipulate neurovascular-specific guidewires and microcatheters. The authors specifically evaluated the movement of the wire forward and backward during the advancement of the microcatheter.RESULTSNavigation of the rete mirabile and an induced aneurysm within the common carotid artery were successful. The active device fixation feature enabled independent advancement and retraction of the guidewire and working device relative to the microcatheter. When ADF was inactive, the mean forward motion of the guidewire was 5 mm and backward motion was 0 mm. When ADF was active, the mean forward motion of the guidewire was 0 mm and backward motion was 1.5 mm. The modifications made to the robotic cassette enabled the system to successfully manipulate the microcatheter and guidewire safely and in a manner more suited to neuroendovascular procedures than before. There were no occurrences of dissection, extravasation, or thrombosis.CONCLUSIONSThe robotic system was originally designed to navigate and manipulate devices for cardiac and peripheral vascular intervention. The current modifications described here improved its utility for the more delicate and tortuous neurovascular environment. This will set the stage for the development of a neurovascular-specific robot.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

Genetics,Animal Science and Zoology

Cited by 41 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. Advances of surgical robotics: image-guided classification and application;National Science Review;2024-06-06

2. Perspectives on Remote Robotic-Assisted Stroke Treatment: A Commentary Paper;American Journal of Neuroradiology;2024-01-12

3. Learning Curve in Robotic Stereoelectroencephalography: Single Platform Experience;World Neurosurgery;2023-11

4. Development of a Slave Manipulator for Vascular Interventional Surgery Robotic System;2023 IEEE International Conference on Mechatronics and Automation (ICMA);2023-08-06

5. Feasibility of robotic neuroendovascular surgery;Interventional Neuroradiology;2023-08-05