Operating room radiation exposure in cone beam computed tomography–based, image-guided spinal surgery

Author:

Nottmeier Eric W.1,Pirris Stephen M.1,Edwards Steven1,Kimes Sherri1,Bowman Cammi1,Nelson Kevin L.2

Affiliation:

1. Departments of Neurosurgery and

2. Radiology, Mayo Clinic, Jacksonville, Florida

Abstract

Object Surgeon and operating room (OR) staff radiation exposure during spinal surgery is a concern, especially with the increasing use of multiplanar fluoroscopy in minimally invasive spinal surgery procedures. Cone beam computed tomography (cbCT)–based, 3D image guidance does not involve the use of active fluoroscopy during instrumentation placement and therefore decreases radiation exposure for the surgeon and OR staff during spinal fusion procedures. However, the radiation scatter of a cbCT device can be similar to that of a standard 64-slice CT scanner and thus could expose the surgeon and OR staff to radiation during image acquisition. The purpose of the present study was to measure radiation exposure at several unshielded locations in the OR when using cbCT in conjunction with 3D image-guided spinal surgery in 25 spinal surgery cases. Methods Five unshielded badge dosimeters were placed at set locations in the OR during 25 spinal surgery cases in which cbCT-based, 3D image guidance was used. The cbCT device (O-ARM) was used in conjunction with the Stealth S7 image-guided platform. The radiology department analyzed the badge dosimeters after completion of the last case. Results Fifty high-definition O-ARM spins were performed in 25 patients for spinal registration and to check instrumentation placement. Image-guided placement of 124 screws from C-2 to the ileum was accomplished without complication. Badge dosimetry analysis revealed minimal radiation exposure for the badges 6 feet from the gantry in the area of the anesthesiology equipment, as well as for the badges located 10–13 feet from the gantry on each side of the room (mean 0.7–3.6 mrem/spin). The greatest radiation exposure occurred on the badge attached to the OR table within the gantry (mean 176.9 mrem/spin), as well as on the control panel adjacent to the gantry (mean 128.0 mrem/spin). Conclusions Radiation scatter from the O-ARM was minimal at various distances outside of and not adjacent to the gantry. Although the average radiation exposure at these locations was low, an earlier study, undertaken in a similar fashion, revealed no radiation exposure when the surgeon stood behind a lead shield. This simple precaution can eliminate the small amount of radiation exposure to OR staff in cases in which the O-ARM is used.

Publisher

Journal of Neurosurgery Publishing Group (JNSPG)

Subject

General Medicine

Cited by 39 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3